CAL 2000LT Complete Instrument
The CAL 2000LT provides unmatched versatility and accuracy in corrosive calibration gases.
Field replaceable electrochemical generating sources provide a gas standard for accurately calibrating chlorine, chlorine dioxide, hydrogen, hydrogen cyanide, and hydrogen sulfide gas sensors.
The CAL 2000LT is microprocessor controlled and has an LCD (display) allowing field adjustable concentration and flow rate. The generating sources are interchangeable and field replaceable. Best of all, the CAL 2000LT generating source provides up to 25times as much calibration gas as a disposable cylinder at approximately the cost of just two cylinders. The CAL 2000LT gas instrument has a built-in mass flow sensor that automatically compensates for altitude and temperature.
CSA NRTL/C Exia T-6 approved intrinsically safe with CE mark and ATEX Certification
Calibration Gas Kit:
This instrument, with one or more required Gas Sources, is a complete calibration gas kit.
Technical Specifications
Available Gas Source(s)
Documentation
Accessories / Replacement Parts
Air Flow Rate (with internal pump) | 0.5 or 1 (flow rate must be specified when ordering) |
Useful Source Life | 10 or 25 hours |
Warm-up time (to 90%) | Approx. 5 minutes. |
L x W x H | 8.50 x 4.25 x 3.00” (21.59 x 10.8 x 7.62 cm) |
Weight | 3 lb. (1360 g) |
Operating Temperature | 0° C to 50° C |
Relative Humidity (intermittent use) | 0 -100% |
Accuracy | ±10% (±5% available) |
Repeatability | ±5% |
Battery Power | 4 alkaline “C” |
Battery Life | 10 hours (at .5 LPM flow rate) |